Drop Counter
Basis of Operation
-
The drop
counter consists of a dispensing reservoir that is positioned above
an optical-path slot. An infrared source sends a beam of
radiation down the slot where it is detected by a photosensitive
sensor. When a drop of solution from the dispenser blocks the beam
of infrared radiation, the data collection program (Logger Pro)
records the
presence of a single drop as a digital event. When another
probe is connected to the LabPro interface, this digital event
signals
Logger Pro to record the probe's current value. By calibrating
the drop counter the number of drops collected can be converted
into a volume.
Auto ID Capable?
- No. Logger Pro must be
informed that the drop counter is connected using the LabPro icon.
Calibration
- In the absence of a calibration,
volume data is recorded in terms of the number of drops counted instead
of a volume. There are two options for calibration:
- Manually enter a drops/mL
ratio based on an earlier calibration or use an approximate value,
such as 20 drops/mL (which is usually accurate to within ±10%)
- Perform a new calibration
using the following steps:
- Fill the reservoir
with deionized water.
- Open the both valves
so that deionized water exits from the reservoir and adjust
the top valve so that a drop is dispensed every couple of
seconds; the time between drops is not critical, provided
that each drop passes through the drop counter before the
succeeding drop. Do not adjust the top valve once it has
been set.
- Close the bottom
valve and add additional deionized water to the reservoir
if needed.
- Weigh a small beaker
and position it so that it can catch the
drops of deionized water after they pass through the Drop
Counter
- Select Experiment:Calibrate:Drop
Counter from the main menu. Begin the calibration by selecting
Calibrate Now; the calibration will begin when the Drop Counter
first detects a drop.
- Open the bottom
valve and allow the calibration software to gather data until
approximately 10 mL of deionized water has been dispensed.
- Determine the mass
of water collected. Measure
the temperature of the water and use its density to calculate
the volume of water collected. Enter this value in the appropriate
space on the software's Calibration page to complete the
calibration.
Preparing Drop Counter for
a pH Titration ( this procedure assumes use of a calibrated pH probe
and a calibrated drop counter)
- Fill the reservoir with
your titrant.
- Open both valves and adjust
the top valve until you have a drop rate of one drop every couple
of seconds; the time between drops is not critical, provided that
each drop passes through the drop counter before the succeeding drop.
Do not adjust the top valve once it has been set.
- Close the bottom valve
and add additional titrant to the reservoir if needed. While using
the drop counter, continue to refill the reservoir as needed.
- Place a stir motor below
the drop counter.
- Place the sample in a suitable
container (usually a small beaker) and place it on the stir motor.
Insert the pH probe and add enough deionized water so that the tip
of the pH probe is
completely
immersed.
- Add either a small stir
bar to the solution or attach the microstirer to the tip of the pH
probe. Adjust the stir rate so that the solution is mixed well.
- Set up your LoggerPro for
data collection, choosing the option for digital collection and for
manually stopping data collection.
- To begin the experiment,
click on LoggerPro's Collect icon. Data collection will begin once
the drop counter senses the presence of the first drop.
- Completely open the lower
valve to allow the titrant to drop into the sample. Data collection
will begin and the resulting graph of pH vs. volume of titrant displayed.
Allow the titration to continue until you are well past the equivalence
point.
- When the titration is complete,
click on LoggerPro's Stop icon to end data collection. Close the
lower valve on the reservoir to stop the flow of titrant to the sample.
Maintenance and Storage
- Rinse the reservoir with
deionized water before storing.
Specifications
- response time is less than
70 ns
- maximum number of drop
is 1504 (approximately 75 mL)
Click here to return to the main page for Using the LabPro Interface.
|